Por favor, use este identificador para citar o enlazar este ítem: https://hdl.handle.net/20.500.12530/54662
Título : A new analytical model for the response curve in megavoltage photon beams of the radiochromic EBT3 films measured with flatbed scanners
Autor : 
Filiación: Hospital Universitario de Fuenlabrada. Servicio de Radiofísica y Protección Radiológica
Palabras clave : 
Mesh: 
Decs: 
Fecha de publicación : 17-may-2022
Editorial : Wiley
Citación : Rodríguez C, García-Pinto D, Martínez LC, López-Fernández A. A new analytical model for the response curve in megavoltage photon beams of the radiochromic EBT3 films measured with flatbed scanners. J Appl Clin Med Phys. 2022;e13654. https://doi.org/10.1002/acm2.13654
Resumen : The aim of this work is to study a new analytical model which describes the dose-response curve in megavoltage photon beams of the radiochromic EBT3 film measured with two commercially available flatbed scanners. This model takes into account the different increase of the number of two types of absorbents in the film with absorbed dose and it allows to identify parameters that depend on the flatbed scanner and the film model, and parameters that exclusively depend on the production lot. In addition, the new model is also compared with other models commonly used in the literature in terms of its performance in reducing systematic calibration uncertainties.
PMID: 35580051
URI : https://hdl.handle.net/20.500.12530/54662
Derechos: info:eu-repo/semantics/openAccess
Atribución-NoComercial-CompartirIgual 3.0 España
ISSN : 1526-9914